Introducing our latest innovation, the Dextran and Sucrose Analysis (DASA) System. The test measures down to 200ppm dextran in solution and easily copes with several thousand, without dilution. This instrument is the latest technology in pol measurements. The SaachAAr 880D is highly accurate (not sensitive to molecular weight), and its one button test generates specific results (distinguishes dextran from other molecules present) including sample preparation in just 10 minutes. The SaachAAr 880D is also environmentally friendly making lead-subacetate redundant. The outputs include: 6 GPL (good laboratory practice) identifiers, Original pol, corrected pol, Dextran ppm, Brix, and Sugar purity.

 
SacchAAr 880

Scales:

I.S.S. (International Sugar Scale 1988 - °Z ), temperature corrected °Z, also angular degrees at 589nm
Range:
Full circle polarimeter
Accuracy:
± 0.01 °Z (0 to 25 °Z), ± 0.001 angular degrees (0 to 10°)
± 0.02 °Z (above 25 °Z), ± 0.01 angular degrees (10 to 90°)
Wavelength:
Sodium Yellow 589.44nm and infrared (IR); Switchable
Readout Time:
Approx. 25 seconds (continuous mode)
Reading Modes:
Continuous. Autoprint - reads continuously, outputs at set times or on data change. One shot and lock - user sets readout time, display locked after time delay. Stability - readout when stable, user sets stability parameters. Dextran, dextran and Purity.
Sample Compartment:
Accepts standard sample tubes with 30mm diameter collars. Maximum path length 200mm, bores 8mm down to 1.5mm. Ventilated - temperature rise above ambient does not exceed 3°C.
Thermometers:
One temperature sensor fitted inside sample compartment plus socket for alternative sample tube temperature sensor. Digital display resolution ± 0.1°C.
Thermostatic Control:
Temperature sensor range 0°C to 100°C, accuracy ± 0.25°C
Controls:
Power ON/OFF, Alpha-numeric keypad, plus function keys including print, zero, and scale selection.
Outputs/Inputs:
2 x RS232 (25-way D sockets); 1 x remote (9-way D socket)
Data Output:
Rotation reading, units, time and date, polarimeter serial number, wavelength, temperature, pre-set parameters, sample number
Additional Sample Identification Feature:
User selection and definition of output parameters for GLP: in addition to the standard data output, up to 6 user defined sample identifiers.
Calibration:
UKAS calibration certificate available if required (order code 10-01)
Power Requirements:
86 to 265 v AC; 47 to 63 Hz , 40 watts maximum
Dimensions
(w x d x h):
515 x 380 x 198 mm
Weight:
15.7kg
Packed Weight:
Approx. 23kg

 
Associated with


3305 Commerce Blvd.
Kissimmee, FL 34741
 

Phone: 407-932-3688
Fax: 407-932-3686
IndexUS@aol.com
 
We Accept Visa and Mastercard
 
*Due to continuing research and technical advances, we reserve the right to alter the product specifications of any of the products on our web site.