Introducing
our latest innovation, the Dextran and Sucrose Analysis (DASA)
System. The test measures down to 200ppm dextran in solution
and easily copes with several thousand, without dilution.
This instrument is the latest technology in pol measurements.
The SaachAAr 880D is highly accurate (not sensitive to molecular
weight), and its one button test generates specific results
(distinguishes dextran from other molecules present) including
sample preparation in just 10 minutes. The SaachAAr 880D is
also environmentally friendly making lead-subacetate redundant.
The outputs include: 6 GPL (good laboratory practice) identifiers,
Original pol, corrected pol, Dextran ppm, Brix, and Sugar
purity.
|
|
| |
SacchAAr
880
|
|
Scales:
|
I.S.S.
(International Sugar Scale 1988 - °Z ), temperature
corrected °Z, also angular degrees at 589nm
|
|
Range:
|
Full
circle polarimeter
|
|
Accuracy:
|
±
0.01 °Z (0 to 25 °Z), ± 0.001 angular
degrees (0 to 10°)
± 0.02 °Z (above 25 °Z), ± 0.01 angular
degrees (10 to 90°)
|
|
Wavelength:
|
Sodium
Yellow 589.44nm and infrared (IR); Switchable
|
|
Readout
Time:
|
Approx.
25 seconds (continuous mode)
|
|
Reading
Modes:
|
Continuous. Autoprint - reads continuously, outputs
at set times or on data change. One shot and lock
- user sets readout time, display locked after
time delay. Stability - readout when stable, user
sets stability parameters. Dextran, dextran and
Purity.
|
|
Sample
Compartment:
|
Accepts
standard sample tubes with 30mm diameter collars.
Maximum path length 200mm, bores 8mm down to 1.5mm.
Ventilated - temperature rise above ambient does
not exceed 3°C. |
|
Thermometers:
|
One temperature sensor fitted inside sample compartment
plus socket for alternative sample tube temperature
sensor. Digital display resolution ± 0.1°C.
|
|
Thermostatic
Control:
|
Temperature sensor range 0°C to 100°C, accuracy
± 0.25°C |
|
Controls:
|
Power ON/OFF, Alpha-numeric keypad, plus function
keys including print, zero, and scale selection. |
|
Outputs/Inputs:
|
2 x RS232 (25-way D sockets); 1 x remote (9-way
D socket) |
|
Data
Output:
|
Rotation
reading, units, time and date, polarimeter serial
number, wavelength, temperature, pre-set parameters,
sample number |
|
Additional
Sample Identification Feature:
|
User selection and definition of output parameters
for GLP: in addition to the standard data output,
up to 6 user defined sample identifiers. |
|
Calibration:
|
UKAS calibration certificate available if required
(order code 10-01)
|
|
Power
Requirements:
|
86 to 265 v AC; 47 to 63 Hz , 40 watts maximum |
|
Dimensions
(w x d x h):
|
515 x 380 x 198 mm |
|
Weight:
|
15.7kg
|
|
Packed
Weight:
|
Approx.
23kg
|
|
|