Model AA-10R with RS232 Output
A new
version of the very popular AA-10 Automatic Polarimeter is
now available with RS232 Output.
Improved
Resolution - now 0.01° Angular or ISS
Optical
Activity quality control laboratory - now UKAS accredited
Externally,
the AA-10R is similar in appearance to the AA-10; internally
the electronics have been modernized to provide a direct RS232
output as standard on all AA-10R's for connection to a standard
printer or computer. Also the LED display digits are larger
and easier to read, and their brightness can be adjusted to
suit personal preference and operating conditions. The reduced
power consumption means the AA-10R polarimeter runs cooler;
the sample compartment is only a few degrees above ambient
temperature. However, for the most accurate polarimetry, we
still recommend the use of jacketed sample tubes; the AA-10R
has a slotted sample compartment cover fitted as standard
which allows thermostatting pipes in and out.
The AA-10R
retains all the excellent features, which have made the AA-10
polarimeter such a successful instrument over the years:
- Robust,
reliable construction - all metal case, with cast aluminum
base frame and solid aluminum panels
-
Small,
compact size - about three-quarters the size of the average
briefcase, important when laboratory bench space is limited
- but still accepts standard size polarimeter tubes up
to 200mm pathlength
- Ease
of use - the AA-10 polarimeter only had 2 controls, an on/off
switch and a zero button - the AA-10R has the same controls
plus a single "print" key to send the results
to a printer (with suitable software, available from Optical
Activity Ltd. the reading can be requested by a computer).
- Proven
accuracy - for all samples including colored and dark samples.
Any sample up to OD1.5 (i.e. a sample, which absorbs less
than 95% of the incident light beam) can be measured accurately.
Biased readings due to colored samples are avoided because
the AA-10 and AA-10R use a halogen lamp in conjunction with
a very narrow band interference filter to select the measuring
wavelength, normally 589.44nm. (The AA-10R is also available
as a dual wavelength model, switchable between 589.44nm
and 546.23nm.)
- A warning
is given if the OD1.5 absorption tolerance is exceeded -
more accurate results can generally be obtained with highly
absorbing or colored samples if the tube length is reduced,
rather than accepting an erroneous reading beyond the OD
tolerance for accurate polarimeter readings. (For full details
of the different sizes and types of sample tube available,
see our Accessories page.
-
Available
either as an ANGULAR scale model with the reading directly
displayed in angular degrees, absolute accuracy +/-0.01°,
OR as a SUGAR SCALE model with the result displayed in
°Z (International Sugar Scale 1988, ISS), absolute
accuracy +/-0.02° Z. Full circle polarimeter, capable
of measuring any optical rotation.
The RS232
data output transmits the AA-10R reading, and, for dual wavelength
models, the wavelength selected. For identification of samples,
a free-standing keyboard is available as an optional extra
which automatically adds time and date to your printed results
and can be used to add alphanumeric identifiers or messages
to each reading. Alternatively, a bar code reader can be used
to label samples.
The Optical
Activity Quality Control Department is a UKAS* accredited
laboratory for the calibration of polarimeters and quartz
control plates. Any Optical Activity polarimeter is available
with a UKAS certificate of calibration if required.
To use
the AA-10R, simply press the set zero button once at the start
of work; pour your solution into a sample tube, place in the
instrument and in a few seconds, the result is shown in the
digital display. Versatility comes from choosing different
tubes for different applications, for example, up to 4 samples
per minute can be measured using a flow cell and either a
funnel or an auto-sampler. The unique operating principle
of the AA-10R, patented by Optical Activity Ltd. in the 1970's,
allows measurement of flowing samples and tolerates some turbulence,
small bubbles or suspended solids which would completely upset
other polarimeters.
This very
versatile polarimeter has so many applications that it is
not possible to list them all here; a few examples are , from
the pharmaceutical industry, products from antibiotics to
cough mixture and dextrose in intravenous fluid. It is used
to determine lactose in milk, starch in animal feed and in
quality control of essential oils where the dual wavelength
model is useful for the colored oils. Throughout the world
sugar industry, the sugar scale version of the AA-10 is used
in both beet and cane sugar factories, tare houses and laboratories.
*UKAS
(The United Kingdom Accreditation Service responsible for
providing national accreditation.)
|
|
|
|
AA-10
Angular
|
 |
AA-10
Sugar
|
|
Scales:
|
Degrees
|
|
ISS
|
|
Range:
|
±
90°
|
|
± 130° Z
|
|
Accuracy:
|
± 0.01%
|
|
±
0.02%Z
|
|
Wavelength:
|
589.44nm
|
|
589.44nm
|
|
Light
Source:
|
Halogen filament lamp, Approx. life 2000 hrs.
|
|
Halogen filament lamp, Approx. life 2000 hrs.
|
|
Outputs:
|
Digital
LED display, plus RS232 25-way output socket
|
|
Digital
LED display, plus RS232 25-way output socket
|
|
Readout
Time:
|
Approx. 8 seconds (for all models, response time
to small changes is one or two seconds)
|
|
Approx.
10 seconds (for all models, response time to small
changes is one or two seconds)
|
|
Sample
Compartment:
|
Accepts standard polarimeter tubes with 30mm circular
ends. Maximum path length 200mm, bores 8mm down
to 1.5mm
|
|
Accepts standard polarimeter tubes with 30mm circular
ends. Maximum path length 200mm, bores 8mm down
to 1.5mm
|
|
Controls:
|
Power
ON/OFF SET ZERO button, PRINT key.
|
|
Power
ON/OFF SET ZERO button, PRINT key.
|
|
Special
Features:
|
Warning if the sample transmittance is less than
the 3% (i.e. OD 15) required for accurate measurement.
|
|
Warning if the sample transmittance is less than
the 3% (i.e. OD 15) required for accurate measurement.
|
|
Power
Requirements:
|
AC
line voltage 230v ± 20% or 115v ± 20%, 45 to 65Hz,
10 watts (please state voltage when ordering)
|
|
AC
line voltage 230v ± 20% or 115v ± 20%, 45 to 65Hz,
10 watts (please state voltage when ordering)
|
|
EMC
and Low Voltage Directive:
|
CE marked complies with the following regulations
89/336/EEC as amended by 91/263/EEC and 92/31/EEC
Electromagnetic Compatibility Directive (EMC)
and 93/68/EEC Low Voltage Directive.
|
|
CE marked complies with the following regulations
89/336/EEC as amended by 91/263/EEC and 92/31/EEC
Electromagnetic Compatibility Directive (EMC)
and 93/68/EEC Low Voltage Directive.
|
|
Dimensions
(w x d x h):
|
460 x 220 x 90mm
|
|
460 x 220 x 90mm
|
|
Weight:
|
8.5kg
|
|
8.5kg
|
|
Packed Weight:
|
12-18kg
|
|
12-18kg
|
|